| Onderwerp (wordt vervangen door Abstract na indienen) | Abstract
This thesis contributed to the further development of the SeaCanairy, a mobile air quality
measurement system. The air circuit was optimized and the electrical diagram
redesigned, making the system more accessible to users with no experience with this
prototype
Significant progress was made with the integration of cabinet heating, which significantly
reduced hygroscopic growth and improved the accuracy of the OPC-N3 sensor.
Furthermore, several technical issues were addressed, such as the influence of the
pump configuration, the need for an additional I²C line, and electromagnetic
interference. Thanks to hardware and software adjustments, the error rate of the OPCN3 sensor decreased from approximately 60% to approximately 3%.
Not all objectives were achieved. For example, the calibration of the Alphasense gas
sensors will need to be repeated in the future, and the flow rate of the OPC-N3 sensor
proved unstable, which noticeably affected the measurement results.
Despite these challenges, this research confirms that the SeaCanairy is a promising
platform for reliable air quality measurements. With further optimizations in calibration,
flow stabilization and software, the system can develop into a robust instrument for longterm measurement campaigns and maritime applications.
|